Lindberg/Blue M LGO 1200°C Box Furnaces
Utilize the latest technical advances in heating elements, insulation and temperature control integrated into a self-contained cabinet with these box furnaces.
Overview
Overview
Utilize the latest technical advances in heating elements, insulation and temperature control integrated into a self-contained cabinet with these box furnaces.
Utilize the latest technical advances in heating elements, insulation and temperature control integrated into a self-contained cabinet with Thermo Scientific™ Lindberg/Blue M™ LGO Box Furnaces. Furnaces feature proprietary LGO (light gauge overbend) heating elements and Thermo Scientific™ Moldatherm™ insulation for efficient and economical transfer of heat to chamber, with low exterior temperatures. Variable heat-up rate eliminates thermal shock to materials with inherently quick heat up and cool down rates.
Key Features
- Choice of side hinge or vertical lift door
- Air vent (1 in. dia., top) and air inlet (0.375 in. dia., rear) for inert atmosphere exchange (may experience some leakage at door)
- Self-tuning, digital instrumentation for precise temperature setpoint and display
- Platinel II™ thermocouple for long-term stability
- 0.6 cu. ft. models feature vertical lift door; 2 cu. ft. models feature horizontal side swing door, hot side facing away from operator for protection
Microprocessor Control
- Microprocessor-based self-tuning PID control (proportional, integral, derivative) provides optimum thermal process, prevents overshoot
- Control panel designed for easy access and maintenance
- Main power ON/OFF switch on control panel
- Adjustable high-limit overtemperature protection
- Simultaneous LED display of actual temperature vs. setpoint
- Can be configured to display temperature in either °C or °F
- Safety door switch interrupts power to heating element when door is opened; protects heating elements and minimizes exposure to end-user
- Removable shelves for versatility
- Moldatherm™ hearthplate supports load and prevents damage due to spillage
Digital, Single-Setpoint Controller
- Single segment, single setpoint, one ramp to setpoint
Digital Single-Program, Multiple-Segment Programmable Controller
- Single program with multiple segments for ramp (up and down) and dwell (timed hold) temperature control
Digital Multiple-Program, Multiple-Segment Programmable Controller
Available on models with (P) suffix designation
- Multiple programs and segments for ramp (up and down) and dwell (timed hold) temperature control
Overtemperature Control (OTC)†
- Protects furnace and load in the event of primary control circuit failure
- Overrides main controller and shuts off power to furnace if high limit is reached
- Manual reset required for safety
- Operates via magnetic contacts through signal from independent thermocouple
Flowmeter Option (FM) (Inert Atmosphere Only)
(FM)
- Gas flowmeter, adjustable, located on front control panel
- Adjustable flowrate, range 1.0 to 10.0 cu. ft./hr. standard
- Suitable for inert gas or airflow to chamber
- Allows fresh air exchange for ashing applications Not suitable for combustible or volatile gases
Ordering Information:
Required power cord and hardwiring not included. Optional RS485 Digital Communications Port allows controller to be connected to a PC for remote monitoring and control of the furnace; up to 30 units can be connected to one PC.
†Specify Option
B for Overtemperature Control (OTC) when ordering.
Includes
One two-part shelf (0.6 cu. ft. models have one shelf position at center position; 2.0 cu. ft. models have three shelf positions)
Lab Furnaces
Table of contents
Lab Box Furnaces
Lab Tube Furnaces
Furnace Accessories
Deliver consistent results with a furnace you can depend on to attain your daily goals. Thermo Scientific Box and Tube Furnaces come in a wide range of offerings to accommodate ordinary and technical tasks alike, such as: temperature ranges up to 1200°C, temperature control options to meet your application needs, and embedded or open heating elements designed to keep samples safe while maintaining reliable temperature uniformity.
Choosing between a box furnace or a tube furnace for your laboratory
Box Furnaces
- Used for processing larger samples or easy access to samples
- Small, medium and large box furnaces available
- Ashing applications in every industry and research laboratory
- Attributes include variable density insulation, double shell cabinets, long-life heating elements and vertical, horizontal side swing or swing down doors
Tube Furnaces
- Used for processing small samples or samples in an inert atmosphere
- Single and three-zoned designs for precise temperature control. Three zone control helps enable different temperatures in each zone.
- Materials testing, pyrolysis, and crystal growing applications
- Some models offer split-hinge design for ease of tube changing
View our quick application guide for choosing a furnace that is suitable for the workload you need. Choose from a variety of temperature control options, heating elements and furnace formats to meet your analytical requirements.
View application guide
Lab Box Furnaces
Typically used for processing larger samples or to easily place and access samples. We offer a versatile selection of small, medium, and large box furnaces suitable for a variety of industrial and laboratory applications. Advanced engineering and specialized construction techniques include variable density insulation, double shell cabinets, long-life heating elements and side-swing doors (vertical or horizontal) or swing down doors.
Lab Tube Furnaces
Tube furnaces are typically used for processing small samples or heating in an inert atmosphere. The single and three-zoned construction is designed to create precise temperature control. Three-zone control enables the user to select a different temperature in each zone (e.g. for gas applications or material experiments). Some models offer split-hinge design, which easily allows you to change the tube.
Furnace Accessories
Accessories for use by furnace
*Additional adapters may be available. Contact your local representative or customer support to order the following: adapters for Lindberg/Blue M 1100 °C furnace model STF55346: 1-in. (Cat. No. 59541TA); adapters for Lindberg/Blue M 1100 °C furnace model STF55666: 6-in. (Cat. No. 59558TA), blank (solid) adapter (Cat. No. 59559TA); and adapters for Lindberg/Blue M 1200 °C split-hinge furnace model HTF55667: 3-in. (Cat. No. 59535TA), 4-in. (Cat. No. 59536TA), 5-in. (Cat. No. 59537TA), 6-in. (Cat. No. 59538TA), blank (solid) adapter (Cat. No. 59539TA).
Replacement Parts
Thermolyne benchtop muffle furnaces
Thermolyne premium large muffle furnaces
Thermolyne largest tabletop muffle furnaces
*Additional replacement parts may be available. Contact your local representative or customer support to order the following: hearth tray for Thermolyne small benchtop muffle furnace model FB1300 (Cat. No. PH44X1); hearth tray for Thermolyne small benchtop muffle furnace model FB1400 (Cat. No. PH48X1); hearth tray for Thermolyne industrial benchtop muffle furnace (Cat. No. PHX2).
Thermo Scientific Furnaces
Furnace Application Guide
Unity Lab Services
Access to our expert service team, information on support plans, parts and consumables resources, and other instrument services
Specifications
Lindberg/Blue M™ LGO Box Furnaces (Cat. No. BF51732C)
| Capacity (English) | 0.6 cu.ft. |
|---|---|
| Capacity (Metric) | 16.4 L |
| Depth (English) Interior | 11 in. |
| Depth (Metric) Interior | 27.9 cm |
| Dimensions (D x W x H) Interior | 11 x 12 x 11 in. (27.9 x 33 x 17.8 cm) |
| Dimensions (L x W x H) Exterior | 23 x 24 x 27 in. (58.4 x 61 x 68.6 cm) |
| Electrical Requirements | 208/240 V 50/60 Hz |
| Frequency | 50/60 Hz |
| Height (English) | 27 in. |
| Height (English) Interior | 11 in. |
| Height (Metric) | 68.6 cm |
| Height (Metric) Interior | 17.8 cm |
| Length (English) | 23 in. |
| Length (Metric) | 58.4 |
| Power Consumption | 4500 w |
| Shipping Weight (Metric) | 75 kg |
| Temperature (Metric) | 100°C to 1200°C |
| Temperature Control | Digital Mutli Seg/1 Prog |
| Type | Multisegment, Single Program |
| Unit Size | Each |
| Voltage | 208/240 V |
| Width (English) | 24 in. |
| Width (Metric) | 61 cm |
| Width (Metric) Interior | 33 cm |
Available models
| Lindberg/Blue M LGO 1200°C Box Furnaces | Lindberg/Blue M Moldatherm Box furnaces | Thermolyne Largest Tabletop Muffle Furnaces | Thermolyne Benchtop Muffle Furnaces | Thermolyne Premium Large Muffle Furnaces | Thermolyne Small Benchtop Muffle Furnaces | |
|---|---|---|---|---|---|---|
| Heat range | 100 - 1200°C | 100 - 1100°C | 100 - 1093°C | 100 - 1200°C | 100 - 1200°C | 100 - 1100°C |
| Available capacity | 16.4 - 55.3 L | 1.99 - 42.5 L | 45 L | 14 L | 2 - 5.8 L | 1.3 to 2.1 L |
| Heating elements | Embedded | Embedded | Embedded | Embedded | Open | Embedded |
Available models
| Lindberg/Blue M 1200°C Split-Hinge Tube Furnaces* | Lindberg/Blue M 1100°C Tube Furnaces | Lindberg/Blue M Mini-Mite Tube Furnaces | |
|---|---|---|---|
| Heat range | 100 - 1200°C | 100 - 1100°C | 100 - 1100°C |
| Available capacity | 30.5, 61, or 91.4 cm / 12, 24, or 36 in. | 61 or 91.4 cm / 24 or 36 in. | 30.5 cm / 12 in. |
| Heating elements | Open | Embedded | Open |
Available models
| Description | Dimensions | Cat. no. | |
|---|---|---|---|
| Thermolyne exhaust tubing kit | |||
| Thermolyne premium large muffle furnaces | Critical to operation - an appropriate exhaust must be installed with your furnace to remove smoke and gases produced during an ashing procedure. Simplify installation when you order this kit. | Kit includes tubing, brackets, and screws. | AY408X1A |
| Thermolyne largest tabletop muffle furnaces | |||
| Lindberg/Blue M tube adapters/sleeves | |||
| Lindberg/Blue M 1100 °C tube furnaces* –STF55346 and STF55666 series models | Increase the flexibility of your applications with additional sized tube adapters. Tube adapters hold process tubes of a certain size in place to ensure the tubes don't contact the sides of the tube furnace, protecting the furnace heating elements. | 3-in. adapter | 59545 |
| 3-in. adapter | 59555 | ||
| 4-in. adapter | 59556 | ||
| Blank (solid) adapter | 59549 | ||
| Lindberg/Blue M 1200 °C split-hinge tube furnaces* – HTF55122, HTF55322, HTF55342, and HTF55347 series models | 0.75-in adapter | 59510 | |
| 1-in. adapter | 59511 | ||
| Blank (solid) adapter | 59519 | ||
| 1-in. adapter | 59521 | ||
| 1.5-in. adapter | 59522 | ||
| 2-in. adapter | 59523 | ||
| 2.5-in. adapter | 59524 | ||
| 3-in. adapter | 59525 | ||
| Blank (solid) adapter | 59529 |
Available models
| Replacement part | For use with | Description | Dimensions | Cat. no. |
|---|---|---|---|---|
| Hearth tray | F47900 Series | Hearth trays and hearth plates sit in the bottom of the furnace and protect the bottom heating element during operation. | 15.2 x 14.3 x 0.95 cm./ 5.98 x 5.63 x 0.37 in. | PH479X1 |
| Hearth tray | F80000 Series | 25.4 x 19.3 x 0.95 cm. / 10 x 7.6 x 0.38 in. | PH480X1 | |
| Ceramic shelf | Ceramic shelves expand interior usable sample space without interfering with max operation temperature. | 17.4 x 17.3 x 1.2 cm. / 6.85 x 6.81 x 0.47 in. | SH480X1 |
Available models
| Replacement part | Description | Dimensions | Cat. no |
|---|---|---|---|
| Crucible Tray | Order extra trays designed to custom configure your furnace. Trays hold crucibles of various sizes, meant to sit on shelves or hearth trays when in use. | 32.4 x 25.4 x 17.1 cm. / 12.75 x 10 x 6.75 in. | TY408X2A |
| Hearth tray | Hearth trays and hearth plates sit in the bottom of the furnace and protect the bottom heating element during operation. | 8.2 x 10.1 x 1.27 cm. / 3.3 x 4 x 0.5 in. | PHX1 |
| Hearth tray | 25.4 x 19.3 x 0.95 cm. / 10 x 7.6 x 0.38 in. | PH177X1 | |
| Stainless steel shelf | Stainless steel shelves expand interior usable sample space, but limit the max operation temperature to 975°C. If higher heat is required, use ceramic shelves. | 32.5 x 21.1 cm. / 12.8 x 8.3 in. | SH408X1 |
| Shelf handle | For use with crucible trays, to pull a hot tray in and out of the furnace. | HN408X2A | |
| Shelf pegs | Holds shelves in place and allows flexibility of shelf placement. | JSX16 |
Available models
| Replacement part | Description | Dimensions | Cat. no |
|---|---|---|---|
| Crucible tray | Order extra trays designed to custom configure your furnace. Trays hold crucibles of various sizes, meant to sit on shelves or hearth trays when in use. | 32.4 x 25.4 x 17.1 cm. / 12.75 x 10 x 6.75 in. | TY408X2A |
| Hearth tray | Hearth trays and hearth plates sit in the bottom of the furnace and protect the bottom heating element during operation. | 17.1 x 14.9 x 1.9 cm. / 6.75 x 5.9 x 0.75 in. | PH146X1 |
| Refractory shelf | Refractory shelves expand interior usable sample space without interfering with max operation temperature. | 35.2 x 25.4 x 1.27 cm. / 13.87 x 10 x 0.56 in. | SH412X1 |
| Shelf handle | For use with crucible trays, to pull a hot tray in and out of the furnace. | HN408X2A |
Documents
Product information, images, documents and videos: Thermo Fisher Scientific. View on manufacturer website Source 2