Thermolyne Premium Large Muffle Furnaces
Conduct industrial applications, including ashing organic and inorganic samples and gravimetric analysis, with this premium large muffle furnace.
Overview
Overview
Conduct industrial applications, including ashing organic and inorganic samples and gravimetric analysis, with this premium large muffle furnace.
Conduct industrial applications including ashing organic and inorganic samples and conducting gravimetric analysis with Thermo Scientific™ Thermolyne™ Premium Large Muffle Furnaces. Unit has four heating elements, which are located on the chamber top, bottom and sides for enhanced temperature uniformity. Built-in vent port removes undesirable contaminants and moisture to extend the life of the element and unit.
- Four heating elements are located on the chamber top, bottom and sides for enhanced temperature uniformity
- Built-in vent port removes undesirable contaminants and moisture to extend the life of the element and unit
- Rear of the chamber incorporates a 0.95cm (0.38in.) dia. port for monitoring chamber temperatures with independent measuring devices
- Optional stainless steel shelf doubles load capacity (maximum temperature of 900°C)
- Door safety switch stops power to heating elements when door opens
- Thermocouple break protection cuts power to the heating elements, preventing a thermocouple failure runaway condition
- Furnaces with B1, C1, and D1 control also use a mechanical overtemperature protection relay
Choice of Temperature Controllers
- Digital single setpoint control (A1): Single display shows actual temperature or setpoint
- Digital single setpoint control with single ramp to setpoint and a dwell (B1): Single display shows actual temperature or setpoint
- Digital programmable control with one stored program of eight segments (C1)
- Digital programmable control with four stored programs, 16 segments per program (D1): Controller includes RS-232 communications
Requires:
All except Models F6010 and F6018 require a conduit connection.
Warranty: 1 year
Disclaimer to internal research use restriction: Any restrictions on the purchaser to utilize the product(s) for internal research purposes only does not apply to the product(s) on the web pages associated with this statement.
Lab Furnaces
Table of contents
Lab Box Furnaces
Lab Tube Furnaces
Furnace Accessories
Deliver consistent results with a furnace you can depend on to attain your daily goals. Thermo Scientific Box and Tube Furnaces come in a wide range of offerings to accommodate ordinary and technical tasks alike, such as: temperature ranges up to 1200°C, temperature control options to meet your application needs, and embedded or open heating elements designed to keep samples safe while maintaining reliable temperature uniformity.
Choosing between a box furnace or a tube furnace for your laboratory
Box Furnaces
- Used for processing larger samples or easy access to samples
- Small, medium and large box furnaces available
- Ashing applications in every industry and research laboratory
- Attributes include variable density insulation, double shell cabinets, long-life heating elements and vertical, horizontal side swing or swing down doors
Tube Furnaces
- Used for processing small samples or samples in an inert atmosphere
- Single and three-zoned designs for precise temperature control. Three zone control helps enable different temperatures in each zone.
- Materials testing, pyrolysis, and crystal growing applications
- Some models offer split-hinge design for ease of tube changing
View our quick application guide for choosing a furnace that is suitable for the workload you need. Choose from a variety of temperature control options, heating elements and furnace formats to meet your analytical requirements.
View application guide
Lab Box Furnaces
Typically used for processing larger samples or to easily place and access samples. We offer a versatile selection of small, medium, and large box furnaces suitable for a variety of industrial and laboratory applications. Advanced engineering and specialized construction techniques include variable density insulation, double shell cabinets, long-life heating elements and side-swing doors (vertical or horizontal) or swing down doors.
Lab Tube Furnaces
Tube furnaces are typically used for processing small samples or heating in an inert atmosphere. The single and three-zoned construction is designed to create precise temperature control. Three-zone control enables the user to select a different temperature in each zone (e.g. for gas applications or material experiments). Some models offer split-hinge design, which easily allows you to change the tube.
Furnace Accessories
Accessories for use by furnace
*Additional adapters may be available. Contact your local representative or customer support to order the following: adapters for Lindberg/Blue M 1100 °C furnace model STF55346: 1-in. (Cat. No. 59541TA); adapters for Lindberg/Blue M 1100 °C furnace model STF55666: 6-in. (Cat. No. 59558TA), blank (solid) adapter (Cat. No. 59559TA); and adapters for Lindberg/Blue M 1200 °C split-hinge furnace model HTF55667: 3-in. (Cat. No. 59535TA), 4-in. (Cat. No. 59536TA), 5-in. (Cat. No. 59537TA), 6-in. (Cat. No. 59538TA), blank (solid) adapter (Cat. No. 59539TA).
Replacement Parts
Thermolyne benchtop muffle furnaces
Thermolyne premium large muffle furnaces
Thermolyne largest tabletop muffle furnaces
*Additional replacement parts may be available. Contact your local representative or customer support to order the following: hearth tray for Thermolyne small benchtop muffle furnace model FB1300 (Cat. No. PH44X1); hearth tray for Thermolyne small benchtop muffle furnace model FB1400 (Cat. No. PH48X1); hearth tray for Thermolyne industrial benchtop muffle furnace (Cat. No. PHX2).
Thermo Scientific Furnaces
Furnace Application Guide
Unity Lab Services
Access to our expert service team, information on support plans, parts and consumables resources, and other instrument services
Specifications
Thermolyne™ Premium Large Muffle Furnaces (Cat. No. F6018)
| Amperage | 11.2 A |
|---|---|
| Capacity (English) | 0.5 cu.ft. |
| Capacity (Metric) | 14 L |
| Certifications/Compliance | CSA approved |
| Depth (English) Interior | 10 in. |
| Depth (Metric) Interior | 25 cm |
| Description | F6018 |
| Dimensions (D x W x H) Interior | 10 x 12.8 x 6.8 in. (25 x 33 x 18 cm) |
| Dimensions (L x W x H) Exterior | 20.1 x 19.1 x 21 in. (51 x 48.5 x 53.3 cm) |
| Electrical Requirements | 208 V 50/60 Hz |
| Frequency | 50/60 Hz |
| Height (English) | 21 in. |
| Height (English) Interior | 6.8 in. |
| Height (Metric) | 53.3 cm |
| Height (Metric) Interior | 18 cm |
| Length (English) | 20.1 in. |
| Length (Metric) | 51 cm |
| Power Consumption | 2325 w |
| Shelves | Stainless Steel |
| Shipping Weight (English) | 134 lb. |
| Shipping Weight (Metric) | 60.8 kg |
| Temperature (Metric) | 100°C to 1200°C |
| Temperature Control | Digital single setpoint (A1) |
| Temperature Stability | ±0.3 at 1000°C |
| Temperature Uniformity | ±2.2 at 1000°C |
| Type | Single Setpoint |
| Unit Size | Each |
| Voltage | 208 V |
| Width (English) | 19.1 in. |
| Width (Metric) | 48.5 cm |
| Width (Metric) Interior | 33 cm |
Available models
| Lindberg/Blue M LGO 1200°C Box Furnaces | Lindberg/Blue M Moldatherm Box furnaces | Thermolyne Largest Tabletop Muffle Furnaces | Thermolyne Benchtop Muffle Furnaces | Thermolyne Premium Large Muffle Furnaces | Thermolyne Small Benchtop Muffle Furnaces | |
|---|---|---|---|---|---|---|
| Heat range | 100 - 1200°C | 100 - 1100°C | 100 - 1093°C | 100 - 1200°C | 100 - 1200°C | 100 - 1100°C |
| Available capacity | 16.4 - 55.3 L | 1.99 - 42.5 L | 45 L | 14 L | 2 - 5.8 L | 1.3 to 2.1 L |
| Heating elements | Embedded | Embedded | Embedded | Embedded | Open | Embedded |
Available models
| Lindberg/Blue M 1200°C Split-Hinge Tube Furnaces* | Lindberg/Blue M 1100°C Tube Furnaces | Lindberg/Blue M Mini-Mite Tube Furnaces | |
|---|---|---|---|
| Heat range | 100 - 1200°C | 100 - 1100°C | 100 - 1100°C |
| Available capacity | 30.5, 61, or 91.4 cm / 12, 24, or 36 in. | 61 or 91.4 cm / 24 or 36 in. | 30.5 cm / 12 in. |
| Heating elements | Open | Embedded | Open |
Available models
| Description | Dimensions | Cat. no. | |
|---|---|---|---|
| Thermolyne exhaust tubing kit | |||
| Thermolyne premium large muffle furnaces | Critical to operation - an appropriate exhaust must be installed with your furnace to remove smoke and gases produced during an ashing procedure. Simplify installation when you order this kit. | Kit includes tubing, brackets, and screws. | AY408X1A |
| Thermolyne largest tabletop muffle furnaces | |||
| Lindberg/Blue M tube adapters/sleeves | |||
| Lindberg/Blue M 1100 °C tube furnaces* –STF55346 and STF55666 series models | Increase the flexibility of your applications with additional sized tube adapters. Tube adapters hold process tubes of a certain size in place to ensure the tubes don't contact the sides of the tube furnace, protecting the furnace heating elements. | 3-in. adapter | 59545 |
| 3-in. adapter | 59555 | ||
| 4-in. adapter | 59556 | ||
| Blank (solid) adapter | 59549 | ||
| Lindberg/Blue M 1200 °C split-hinge tube furnaces* – HTF55122, HTF55322, HTF55342, and HTF55347 series models | 0.75-in adapter | 59510 | |
| 1-in. adapter | 59511 | ||
| Blank (solid) adapter | 59519 | ||
| 1-in. adapter | 59521 | ||
| 1.5-in. adapter | 59522 | ||
| 2-in. adapter | 59523 | ||
| 2.5-in. adapter | 59524 | ||
| 3-in. adapter | 59525 | ||
| Blank (solid) adapter | 59529 |
Available models
| Replacement part | For use with | Description | Dimensions | Cat. no. |
|---|---|---|---|---|
| Hearth tray | F47900 Series | Hearth trays and hearth plates sit in the bottom of the furnace and protect the bottom heating element during operation. | 15.2 x 14.3 x 0.95 cm./ 5.98 x 5.63 x 0.37 in. | PH479X1 |
| Hearth tray | F80000 Series | 25.4 x 19.3 x 0.95 cm. / 10 x 7.6 x 0.38 in. | PH480X1 | |
| Ceramic shelf | Ceramic shelves expand interior usable sample space without interfering with max operation temperature. | 17.4 x 17.3 x 1.2 cm. / 6.85 x 6.81 x 0.47 in. | SH480X1 |
Available models
| Replacement part | Description | Dimensions | Cat. no |
|---|---|---|---|
| Crucible Tray | Order extra trays designed to custom configure your furnace. Trays hold crucibles of various sizes, meant to sit on shelves or hearth trays when in use. | 32.4 x 25.4 x 17.1 cm. / 12.75 x 10 x 6.75 in. | TY408X2A |
| Hearth tray | Hearth trays and hearth plates sit in the bottom of the furnace and protect the bottom heating element during operation. | 8.2 x 10.1 x 1.27 cm. / 3.3 x 4 x 0.5 in. | PHX1 |
| Hearth tray | 25.4 x 19.3 x 0.95 cm. / 10 x 7.6 x 0.38 in. | PH177X1 | |
| Stainless steel shelf | Stainless steel shelves expand interior usable sample space, but limit the max operation temperature to 975°C. If higher heat is required, use ceramic shelves. | 32.5 x 21.1 cm. / 12.8 x 8.3 in. | SH408X1 |
| Shelf handle | For use with crucible trays, to pull a hot tray in and out of the furnace. | HN408X2A | |
| Shelf pegs | Holds shelves in place and allows flexibility of shelf placement. | JSX16 |
Available models
| Replacement part | Description | Dimensions | Cat. no |
|---|---|---|---|
| Crucible tray | Order extra trays designed to custom configure your furnace. Trays hold crucibles of various sizes, meant to sit on shelves or hearth trays when in use. | 32.4 x 25.4 x 17.1 cm. / 12.75 x 10 x 6.75 in. | TY408X2A |
| Hearth tray | Hearth trays and hearth plates sit in the bottom of the furnace and protect the bottom heating element during operation. | 17.1 x 14.9 x 1.9 cm. / 6.75 x 5.9 x 0.75 in. | PH146X1 |
| Refractory shelf | Refractory shelves expand interior usable sample space without interfering with max operation temperature. | 35.2 x 25.4 x 1.27 cm. / 13.87 x 10 x 0.56 in. | SH412X1 |
| Shelf handle | For use with crucible trays, to pull a hot tray in and out of the furnace. | HN408X2A |
Documents
Product information, images, documents and videos: Thermo Fisher Scientific. View on manufacturer website Source 2